Transport, scope of delivery, installation
14
Ambient temperature
– Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.
• Do not exceed or fall below the permitted temperature range of the
machine (5°C to 40°C / ambient temperature).
Pos: 7.13 /00003 Standard Kapitel/ General Modul Luftfeuchtigkeit @ 0\mod_ 12289185386 93_9.doc @ 4 738 @ @ 1
Atmospheric humidity:
Maximum relative humidity 80% at temperatures up to 31°C, decreasing linearly up
to 50% relative humidity at 40°C
Pos: 7.14 /00004 Warnhinweis e/H0011 HINW EIS Luftfeuchtig keit @ 0\mod _122891888262 8_9.doc @ 478 8 @ @ 1
Atmospheric humidity
– Electronic and mechanical components may be damaged and the
performance data alter to an unknown extent.
• Do not exceed the admissible range for atmospheric humidity.
Pos: 7.15 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 A ufstellen des Ger ätes @ 0\ mod_12264988 49756_9.doc @ 3464 @ 2 @ 1
4.5 Installation of the machine
Pos: 7.16 /00010 Bedienungsa nleitungen Kapitels ammlungen/AS 450_basic/00 06 Transport, Li eferumfang un d Aufstellen /AS 450 basic Mod ul Gerät a usrichten @ 4\ mod_1329993637 090_9.doc @ 26800 @ @ 1
The device should be erected on a level and stable surface which will not sag and
which is suitable for use during operation.
A level surface ensures the even distribution of the sample across the sieves and
stability of the machine.
Fig. 1: Aligning the device
The flexible feet underneath the machine enables it to be adjusted to the surface.
• Unscrew the lock nut (KM).
• Twist the flexible foot of the device (GF) out or in as necessary.
• Tighten the lock nuts on the feet.
Pos: 7.17 /00003 Standard Kapitel/ General Modul Aufstellungsh öhe @ 0\m od_1228918538 349_9.doc @ 4 724 @ @ 1
Installation height: maximum 2000 m above sea level
Pos: 7.18 /00004 Warnhinweis e/H0058 HINW EIS Boden Aufst ellen Vibration en Boden (AS45 0 basic) @ 4\ mod_133294 3698568_9.doc @ 27538 @ @ 1
Erection
– Vibrations can occur depending on the operating status of the device.
• Only erect the device on a level, flat and horizontal surface. The base
must be stable and must not vibrate.
Pos: 7.19 /00005 Ü berschriften/ 1.1 Überschrifte n/1.1 Überschrift en BDA/11 Sp annstange einset zen @ 4\ mod_1330075289 191_9.doc @ 26818 @ 2 @ 1