Cleaning and service
32
The AS450 basic is protected by a device fuse .
The external fusing must be T5.0A (230V) F15A (110V).
Turn the device off at the mains (H).
Pull the mains plug from the low heat device socket (J).
Unscrew the fuse insert (I) and remove the fuse (IS).
CAUTION
A fuse must be replaced by one of the same type and with the same fuse value.
Place the new fuse in the fuse insert.
NOTE
Do not tighten the fuse insert too much when screwing the fuse in!
Screw the fuse insert (I) back in.
NOTE
If the fuse triggers again immediately after being replaced, there is a fault in the
device which must be rectified.
There are two fuses: one in the live conductor and one in the neutral conductor of
the power supply.