Operating the machine
26
The third digit for tenths of a minute is displayed in the operating time (At)
display.
Press the + or - (B3) / (B4) buttons to set the sieve time in tenths of a
minute.
Display in 1/10 of
a minute
0
1
2
3
4
5
6
7
8
9
Resultant value in
seconds
0
6
12
18
24
30
36
42
48
54
5.9.3 Amplitude
Press the SETTING button a third time.
The amplitude value is displayed in the amplitude (Aa) display.
Press the + or - (B3) / (B4) buttons to set the amplitude.
The settings “0” to “9” are proportional to the maximum value of 1.5 mm at 50 Hz
and 2.0 mm at 60 Hz.
5.9.4 Interval sieve time (interval duration)
The cycle time can be set to a value between 1 and 99 seconds.
Press the SETTING button a fourth time.
The interval duration is displayed in seconds in the operating time (At)
display.
The following symbol appears in the amplitude (Aa) display at the top and
bottom alternately: